TY - JOUR
T1 - A novel wireless implant for central venous pressure measurement
T2 - First animal experience
AU - Manavi, Tejaswini
AU - Vazquez, Patricia
AU - O'Grady, Helen
AU - Martina, Jerson
AU - Rose, Michael
AU - Nielsen, Douglas
AU - Fitzpatrick, David
AU - Forouzan, Omid
AU - Nagy, Michael
AU - Sharif, Faisal
AU - Zafar, Haroon
N1 - Publisher Copyright:
© 2020 Heart Rhythm Society
PY - 2020/11/1
Y1 - 2020/11/1
N2 - Background/Objective: Central venous pressure (CVP) serves as a surrogate for right atrial pressure, and thus could potentially predict a wider range of heart failure conditions. The purpose of this work is to assess CVP, through an implantable sensor incorporated with a novel anchor design, in the inferior and superior vena cava of an animal model. Methods: Two animals (Dorset sheep) were implanted with sensors at 3 different locations: inferior vena cava (IVC), superior vena cava (SVC), and pulmonary artery (PA). Two sensors with distinct anchor designs considering anatomical requirements were used. A standard PA sensor (trade name Cordella) was deployed in the PA and SVC, whereas a sensor with a modified cylindrical anchor with various struts was designed to reside in the IVC. Each implant was calibrated against a Millar catheter reference sensor. The ability of the central venous sensors to detect changes in pressure was evaluated by modifying the fluid volume of the animal. Results: The sensors implanted in both sheep were successful, which provided an opportunity to understand the relationship between PA and CVP via simultaneous readings. The mapping and implantation in the IVC took less than 15 minutes. Multiple readings were taken at each implant location using a hand-held reader device under various conditions. CVP recorded in the IVC (6.49 mm Hg) and SVC (6.14 mm Hg) were nearly the same. PA pressure (13–14 mm Hg) measured was higher than CVP, as expected. The SVC waveforms showed clear beats and respiration. Respiration could be seen in the IVC waveforms, but not all beats were easily distinguishable. Both SVC and IVC readings showed increases in pressure (3.7 and 2.7 mm Hg for SVC and IVC, respectively) after fluid overload was induced via extra saline administration. Conclusion: In this work, the feasibility of measuring CVP noninvasively was demonstrated. The established ability of wireless PA pressure sensors to enable prevention of decompensation events weeks ahead can now be explored using central venous versions of such sensors.
AB - Background/Objective: Central venous pressure (CVP) serves as a surrogate for right atrial pressure, and thus could potentially predict a wider range of heart failure conditions. The purpose of this work is to assess CVP, through an implantable sensor incorporated with a novel anchor design, in the inferior and superior vena cava of an animal model. Methods: Two animals (Dorset sheep) were implanted with sensors at 3 different locations: inferior vena cava (IVC), superior vena cava (SVC), and pulmonary artery (PA). Two sensors with distinct anchor designs considering anatomical requirements were used. A standard PA sensor (trade name Cordella) was deployed in the PA and SVC, whereas a sensor with a modified cylindrical anchor with various struts was designed to reside in the IVC. Each implant was calibrated against a Millar catheter reference sensor. The ability of the central venous sensors to detect changes in pressure was evaluated by modifying the fluid volume of the animal. Results: The sensors implanted in both sheep were successful, which provided an opportunity to understand the relationship between PA and CVP via simultaneous readings. The mapping and implantation in the IVC took less than 15 minutes. Multiple readings were taken at each implant location using a hand-held reader device under various conditions. CVP recorded in the IVC (6.49 mm Hg) and SVC (6.14 mm Hg) were nearly the same. PA pressure (13–14 mm Hg) measured was higher than CVP, as expected. The SVC waveforms showed clear beats and respiration. Respiration could be seen in the IVC waveforms, but not all beats were easily distinguishable. Both SVC and IVC readings showed increases in pressure (3.7 and 2.7 mm Hg for SVC and IVC, respectively) after fluid overload was induced via extra saline administration. Conclusion: In this work, the feasibility of measuring CVP noninvasively was demonstrated. The established ability of wireless PA pressure sensors to enable prevention of decompensation events weeks ahead can now be explored using central venous versions of such sensors.
KW - Central venous pressure
KW - Heart failure
KW - Hemodynamic sensor
KW - Wireless hemodynamic monitoring
KW - Wireless implant
UR - http://www.scopus.com/inward/record.url?scp=85139257812&partnerID=8YFLogxK
U2 - 10.1016/j.cvdhj.2020.10.004
DO - 10.1016/j.cvdhj.2020.10.004
M3 - Article
AN - SCOPUS:85139257812
SN - 2666-6936
VL - 1
SP - 130
EP - 138
JO - Cardiovascular Digital Health Journal
JF - Cardiovascular Digital Health Journal
IS - 3
ER -