TY - JOUR
T1 - Establishment of an Automated Algorithm Utilizing Optical Coherence Tomography and Micro-Computed Tomography Imaging to Reconstruct the 3-D Deformed Stent Geometry
AU - Elliott, Mark R.
AU - Kim, Dan
AU - Molony, David S.
AU - Morris, Liam
AU - Samady, Habib
AU - Joshi, Sarang
AU - Timmins, Lucas H.
N1 - Publisher Copyright:
© 1982-2012 IEEE.
PY - 2019/3
Y1 - 2019/3
N2 - Percutaneous coronary intervention (PCI) is the prevalent treatment for coronary artery disease, with hundreds of thousands of stents implanted annually. Computational studies have demonstrated the role of biomechanics in the failure of vascular stents, but clinical studies is this area are limited by a lack of understanding of the deployed stent geometry, which is required to accurately model and predict the stent-induced in vivo biomechanical environment. Herein, we present an automated method to reconstruct the 3-D deployed stent configuration through the fusion of optical coherence tomography (OCT) and micro-computed tomography ( μ CT) imaging data. In an experimental setup, OCT and μ CT data were collected in stents deployed in arterial phantoms (n=4). A constrained iterative deformation process directed by diffeomorphic metric mapping was developed to deform μ CT data of a stent wireframe to the OCT-derived sparse point cloud of the deployed stent. Reconstructions of the deployed stents showed excellent agreement with the ground-truth configurations, with the distance between corresponding points on the reconstructed and ground-truth configurations of 184 ± 96μm. Finally, reconstructions required <30 min of computational time. In conclusion, the developed and validated reconstruction algorithm provides a complete spatially resolved reconstruction of a deployed vascular stent from commercially available imaging modalities and has the potential, with further development, to provide more accurate computational models to evaluate the in vivo post-stent mechanical environment, as well as clinical visualization of the 3-D stent geometry immediately following PCI.
AB - Percutaneous coronary intervention (PCI) is the prevalent treatment for coronary artery disease, with hundreds of thousands of stents implanted annually. Computational studies have demonstrated the role of biomechanics in the failure of vascular stents, but clinical studies is this area are limited by a lack of understanding of the deployed stent geometry, which is required to accurately model and predict the stent-induced in vivo biomechanical environment. Herein, we present an automated method to reconstruct the 3-D deployed stent configuration through the fusion of optical coherence tomography (OCT) and micro-computed tomography ( μ CT) imaging data. In an experimental setup, OCT and μ CT data were collected in stents deployed in arterial phantoms (n=4). A constrained iterative deformation process directed by diffeomorphic metric mapping was developed to deform μ CT data of a stent wireframe to the OCT-derived sparse point cloud of the deployed stent. Reconstructions of the deployed stents showed excellent agreement with the ground-truth configurations, with the distance between corresponding points on the reconstructed and ground-truth configurations of 184 ± 96μm. Finally, reconstructions required <30 min of computational time. In conclusion, the developed and validated reconstruction algorithm provides a complete spatially resolved reconstruction of a deployed vascular stent from commercially available imaging modalities and has the potential, with further development, to provide more accurate computational models to evaluate the in vivo post-stent mechanical environment, as well as clinical visualization of the 3-D stent geometry immediately following PCI.
KW - Optical coherence tomography
KW - biomechanics
KW - coronary artery disease
KW - diffeomorphic metric mapping
KW - vascular stent
UR - http://www.scopus.com/inward/record.url?scp=85053291250&partnerID=8YFLogxK
U2 - 10.1109/TMI.2018.2870714
DO - 10.1109/TMI.2018.2870714
M3 - Article
C2 - 30843790
AN - SCOPUS:85053291250
SN - 0278-0062
VL - 38
SP - 710
EP - 720
JO - IEEE Transactions on Medical Imaging
JF - IEEE Transactions on Medical Imaging
IS - 3
M1 - 8466919
ER -