T1 - Insights from complex aortic surgery with a Streamliner device for aortic arch repair (STAR)
AU - Stefanov, Florian
AU - Morris, Liam
AU - Elhelali, Ala
AU - Kavanagh, Edel P.
AU - Lundon, Violet
AU - Hynes, Niamh
AU - Sultan, Sherif
N1 - Publisher Copyright:
© 2016 The American Association for Thoracic Surgery
PY - 2016/11/1
Y1 - 2016/11/1
N2 - Objective Aortic arch aneurysms and thoracoabdominal aortic arch aneurysms are technically challenging to manage by established surgical and endovascular methods. The Streamliner Multilayer Flow Modulator device (Cardiatis, Isnes, Brussels, Belgium) offers an unorthodox option for these high-risk cases. The Streamliner device for aortic arch repair (STAR) study investigated complex aneurysm cases managed by the Streamliner Multilayer Flow Modulator device and offers an analytic solution for a clinical dilemma. Methods Six cases were included, with a 1-year follow-up, comprising 4 pure arch aneurysms and 2 thoracoabdominal aortic arch aneurysms Crawford type I, from a multicenter database hosted by the Streamliner Multilayer Flow Modulator Global Registry. A total of 50% of cases were performed under instructions for use. All were American Society of Anesthesiology IV and originated from zone 0. All cases were computationally analyzed, which consisted of (1) simulating the treatment on the basis of the postoperative data, (2) repositioning the stents for the failed technical cases, and (3) assessing the effects of overlapping devices on branch patency. Results Correct device placement induced aneurysm flow streamlining, which reduced the dynamic pressure by 23% to 66%, whereas incorrect placements promoted Failure Mode I with 58% and 16% dynamic pressure increases and aneurysm volume expansion up to 23%. Overlapped devices improved distal perfusion by increasing arch branch outflows from 5% to 24%. The Streamliner Multilayer Flow Modulator device does not benefit a sac volume greater than 400 cm3. Conclusions The Streamliner Multilayer Flow Modulator device is a new technology that can manage complex aortic arch aneurysms and thoracoabdominal aortic arch aneurysms with favorable clinical outcomes if it is performed under instructions for use. Careful procedure planning and perioperative virtual stent placement will avoid foreshortening, prevent inadequate stent overlap lengths, and provide insight into the sufficient numbers of required implanted devices.
AB - Objective Aortic arch aneurysms and thoracoabdominal aortic arch aneurysms are technically challenging to manage by established surgical and endovascular methods. The Streamliner Multilayer Flow Modulator device (Cardiatis, Isnes, Brussels, Belgium) offers an unorthodox option for these high-risk cases. The Streamliner device for aortic arch repair (STAR) study investigated complex aneurysm cases managed by the Streamliner Multilayer Flow Modulator device and offers an analytic solution for a clinical dilemma. Methods Six cases were included, with a 1-year follow-up, comprising 4 pure arch aneurysms and 2 thoracoabdominal aortic arch aneurysms Crawford type I, from a multicenter database hosted by the Streamliner Multilayer Flow Modulator Global Registry. A total of 50% of cases were performed under instructions for use. All were American Society of Anesthesiology IV and originated from zone 0. All cases were computationally analyzed, which consisted of (1) simulating the treatment on the basis of the postoperative data, (2) repositioning the stents for the failed technical cases, and (3) assessing the effects of overlapping devices on branch patency. Results Correct device placement induced aneurysm flow streamlining, which reduced the dynamic pressure by 23% to 66%, whereas incorrect placements promoted Failure Mode I with 58% and 16% dynamic pressure increases and aneurysm volume expansion up to 23%. Overlapped devices improved distal perfusion by increasing arch branch outflows from 5% to 24%. The Streamliner Multilayer Flow Modulator device does not benefit a sac volume greater than 400 cm3. Conclusions The Streamliner Multilayer Flow Modulator device is a new technology that can manage complex aortic arch aneurysms and thoracoabdominal aortic arch aneurysms with favorable clinical outcomes if it is performed under instructions for use. Careful procedure planning and perioperative virtual stent placement will avoid foreshortening, prevent inadequate stent overlap lengths, and provide insight into the sufficient numbers of required implanted devices.
KW - aneurysm
KW - aortic arch
KW - endovascular
KW - multilayer flow modulator
KW - streamliner
KW - volume
UR - http://www.scopus.com/inward/record.url?scp=84992359680&partnerID=8YFLogxK
U2 - 10.1016/j.jtcvs.2016.06.043
DO - 10.1016/j.jtcvs.2016.06.043
M3 - Article
C2 - 27485677
AN - SCOPUS:84992359680
SN - 0022-5223
VL - 152
SP - 1309-1318.e5
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
IS - 5
ER -