Abstract
We present a programmable array microscope that uses a ferroelectric liquid-crystal spatial light modulator (SLM) for dynamic generation of scanning apertures. A single SLM serves as both the source and the detector aperture array in a double-pass confocal system. Successive aperture frames scan the array across the viewing area for complete imaging of a sample while preserving depth discrimination. Integration of the microscope output across all aperture frames produces a confocal image.
| Original language | English |
|---|---|
| Pages (from-to) | 2664-2669 |
| Number of pages | 6 |
| Journal | Applied Optics |
| DOIs | |
| Publication status | Published - 1 Jun 2000 |
| Externally published | Yes |
Keywords
- ferroelectric materials
- liquid crystals
- microscopic examination
- scanning