Abstract
Computer-Assisted surgery is becoming more frequently used in the medical world. OrthoPilot™ Hip Suite CT-free navigation system (BBraun Aesculap, Tuttlingen, Germany) is one such computer-Assisted navigation system used for total hip replacement surgery. The validity of OrthoPilot system remains to be determined independent of the manufacturer. The main aims of this study were to investigate the accuracy of OrthoPilot data while using surgical instruments and to confirm the suitability of the cup navigation algorithm. OrthoPilot was compared with the gold standard of a VICON movement analysis system. An aluminium pelvic phantom was machined with high accuracy to perform the experiments. Data were captured simultaneously from both OrthoPilot and VICON systems and acetabular angles were compared. Both systems produced comparable results for the distance between anatomical landmarks and acetabular angles. It can be concluded that data from the OrthoPilot system, if used correctly, are accurate enough for orthopaedic applications.
Original language | English |
---|---|
Pages (from-to) | 315-329 |
Number of pages | 15 |
Journal | International Journal of Biomedical Engineering and Technology |
Volume | 17 |
Issue number | 4 |
DOIs | |
Publication status | Published - 2015 |
Externally published | Yes |
Keywords
- Anterior pelvic plane
- Anteversion angle
- Computer-Assisted surgery
- Cup navigation
- Hip prosthesis
- Inclination angle
- Surgical systems
- Validation